The output "Temperatures" cluster use an interger format for both thermocouple and cold junction readings. It should use double, just like the input bundle did to form cluster. What happens is temperature reading looses resolution. should be 0.25C resolution but sofware makes it 1C.
The output "Temperatures" cluster use an interger format for both thermocouple and cold junction readings. It should use double, just like the input bundle did to form cluster. What happens is temperature reading looses resolution. should be 0.25C resolution but sofware makes it 1C.
See https://forum.digilentinc.com/topic/15498-software-error-pmodtc1lvlibreadvi/