estvar always returns 0.00000 #205
Comments
I get some (not very accurate) result using simulated data: @sidward Can you take a look? Also: Isn't there a formula for predicting the eigenvalues of the noise calibration matrix instead of doing simulations? |
Hi all, @jonasteuwen Do you mind linking me to the fast mri dataset so that I can take a look? @uecker I used simulations as, after the initial simulation, it's quick to re-load and use the simulation results. I do not know how to calculate the expected noise distribution of a matrix with block-hankel structure. Instead, I can look into projecting the low-res image onto an ortho-normal image-sparse basis and use the low-amplitude coefficients to estimate the variance. Would that work? |
@sidward I like the approach in principle. I also do not know an analytical formula, but it sounds like a problem where one might find results in the literature. Another approach might be to use the calibration matrix to project onto the noise subspace and then go back to k-space. Then read off the noise level (correcting for the size of the subspace). |
@uecker I'll look into the literature, but it will be a while before I can do it justice. As for the noise-subspace projection, i'll look into it. I think in principle, it is equivalent to solving an ACS-limited ESPIRiT problem, and subtracting the ACS data from the solution (after projecting the result back to the coil subspace). That being said, this too will take some time for me to get to unfortunately as I will be moving soon. |
@sidward You need do download the dataset here: fastmri.org/dataset I cannot share it myself. |
@sidward no problem. |
Sorry, I forgot to respond: Yes, the dimensions are correct. The tool expects the first three dimensions to be spatial and the fourth dimension to be coil. |
As input I have data of shape (num_slices, height, width, num_coils) made by
writecfl
(coming from the fastmri dataset). However, when I runbart estvar -k1 -r <number_of_central_lines>
I always getEstimated noise variance: 0.00000