Releases: GenyEEM/Updates
EEM calibration software fixed up and enhanced
Enhancements and fixes to EEM sequence test programme
-
Optimised sequence test programme for the accuracy test / dial test
-
Added a time base option to the dial test and optimised the test.
-
Added display of the number of impulses for both the creep test and the start test.
-
Minor changes to the interface of the query program and optimisation of the display of the results
-
The reset timing of ICT equipment is set to 5 seconds after the current is reset to zero.
-
Enhance the program's query data interface to improve usability, and ensure it generates a folder named "c:\temp_EEM" for exporting the Excel file.
-
The query and print program prioritizes using WPS to open Excel files.
-
Fixed issue with #the Neutral switching not working properly when connecting G-480 device through a router, and found a wait delay issue
-
A no-load test can be adjusted to a specific percentage of the voltage.
-
The issue causing prolonged delays in retrieving error data for Type-D benches has been resolved by updating the DLL file, prjErrCalMgr.dll.
-
Resolved the issue with the switching of the neutral line.
EEM Software Database Synchronization Tool
-
First beta version of SQL server server data synchronization
latest upgraded file version
MeterTest.exe, file version: 1.4.371
DataUploadTool, file version 1.2
query test data & print out, file version 1.1.49
Toolkit program, 1.1.4.37
DeviceControl.dll, 1.1.2.116
prjErrCalMgr.dll, 1.0.0.44
EEM calibration software fixed up and enhanced
Improved and fixed up meter test program incl. toolkit
- Fixed an exception due to an incorrect reference standard meter constant in the Dial Test.
- Fixed issue # Switching current loop for D-Test Bench
- Added new Excel template for exporting test data
- Optimised sequence test procedure and fixed some bugs.
- Fixed several errors with API functions in the DLL
- Optimisation of the dial test for the G-type test bench
latest upgraded file version
MeterTest.exe, file version: 1.4.326
Yc189d_toolkit.exe, file version: 1.1.4.37
DeviceControl.dll, version: 1.1.2.115
1.4.42 Upgrade Package Program
EEM calibration software fixed up and enhanced
Improved and fixed up meter test program incl. toolkit
-
Fixed and optimized multi-serial parallel communication
-
Add the database function to the toolkit program incl. support SQL database etc., it can query and export the test data, the function will be improved continuously
-
Fixed issue with #Displaying Repeatability Test Data while browsing
-
Fixed issue #2p2w/ab output for special wiring method
-
fixed issue 1P3W output at DeviceControl.dll
-
fixed issue about Radian RX31 communicaion at sync of BNC / readings
-
fixed issue at EEM's register programme for the meter under test
-
Up to 20 nos of current points can be set at will
-
Right-clicking on the load point table when creating the calibration scheme brings up an editing menu
-
The time of the sequence tests is accumulated by the programme.
-
Add export functionality for plain text files such as TXT, PDF, and XML
-
The G-type test bench now features enhanced functionality with automatic switching between 2 and 7 test current loops, along with updated DLL releases.
EEM Cal for sub standard meter (ver 1.1)
- The new installer can be downloaded with file name setupNewspace.exe
latest upgraded file version
MeterTest.exe, file version: 1.4.296
Yc189d_toolkit.exe, file version: 1.1.4.33
Vector.exe, file version:1.0.0.7
DeviceControl.dll, file version 1.1.2.114
Ctrcomm.dll, file version 2.2.47
InputNo.exe , file version 1.0.2
EEM calibration software fixed up and enhanced
Improved and fixed up meter test program incl. toolkit
-
Added new function of fine tuning for 3-phase output
-
Change the length of the two database fields serialno/meterno to 50 characters.
-
Optimize the performance error readings for 48-postion test bench
use "/" as separator of error calculators grouping to separate the position sequence, such like 1,2,3,4/5,6,7,8/.../45,46,47,48 -
added new feature finetune of manual input in order to manually calibrate the power source when the standard meter is not available. The program is completed according to the following steps
4.1 Output according to parameters such as rated voltage and current, as shown as pic as below
4.2 Switch to the fine-tune output page and complete the manual calibration on the right side of the screen.
4.3 button to perform a fine-tuning operation, the correction value will be displayed in the left input box, the next operator to observe the standard meter readings, entered into the right input box, and then perform another fine-tuning, and so on to know that the accuracy meets the requirements.
Note that the fine-tuning values are stored in the file and will be read back next time and displayed in the left input box.
-
The metertest program of EEM software has added the new finetune feature
how to open the fine-tune of power source
5.1 select the fine-tune option in system setup
5.2 Select the "fine-tune" option in the test project.
5.3 The power source can be adjusted to enhance accuracy during testing.
-
Toolkit program added the function to read the instantaneous test data of Radian RD meter, and fine-tune supports RD meter as well.
-
Repeatability test can be done in four quadrants.
-
Supports real-time temperature readings from thermohygrometer model TH11S
-
The issue with the standard K8 meter's three-wire mode control and the Dll by name ctrcomm.dll has been resolved.
-
improved dail test performance.
-
Toolkit program with new feature about Relay test for three-phase test bench with ICTs
-
the EEM sequence test program added new option to reset the ICTs while current goes zero for test bench with ICTs.
-
Test data query program optimizes text output formatting.
latest upgraded file version
MeterTest.exe, file version: 1.4.279
Yc189d_toolkit.exe, file version: 1.1.3.29
DeviceControl.dll, file version:1.1.2.106
DataManage.exe, file version: 1.1.40
Ctrcomm.dll, file version 2.2.44
EEM calibration software fixed up and enhanced
Improved and fixed up meter test program, and the latest version number is up to 1.4.231
- Optimized simultaneous operation of multiple test benches
- Optimized error readings
- Fixed dial test issue with portable test system YCSS101
- Optimized variation frequency control for hybrid test bench
- Fixed issue #frequency variation test operates in three-phase only in hybrid bench
Upgrade package program version up to 1.4.39 updated on 2023-06-18
The file GENY.Software.Update.1.4.39.EXE contains the following updated files
Metertest.exe, file version 1.4.228
prjErrCalMgr.dll, file version 1.0.0.43
EEM calibration software fixed up and enhanced
improved and fixed up toolkit program with the latest version number 1.1.3.22
- added reactive cross-connected measurement control for reference meter K6/K8
- Q-1 Reactive power
- Q3-3 Reactive power
- Q3-3 Reactive (Cross-connected)
- Q3-2 Reactive power
- Q3-2 Reactive (Cross-connected)
- Added Gating Test feature and gating system is required
- Added communication test for Eletechsup 2 Channel Serial Port Relay Module
- Dial test added a new check for cutting off the current loop, closing the relay before the test, and disconnecting it after the test end,
Fixed issues for API library "Ctrcomm.dll" with file version number 2.2.43
- added reactive cross-connected measurement control
Improved and fixed up meter test program, and the latest version number is up to 1.4.205
- Add a confirmation prompt box before starting the test
- fixed up the issue about the program not response in manual mode
- fixed up the issue about Warmup for test mode 1-phase 3-wire / 2-phase three wire
- fixed up the issue compatible with the old count device for the Dial test
- fixed up the issue of reactive power for the 2-phase 3-wire (/AB) wire mode
- Monitor add new readings for apparent power (VA)
- The dial test allows to set the number of repetitions, the test can be repeated N times, and the results are recorded
- Added option for voltage/current precise output; accuracy can be set up to 0.01% by fine-tuning within 2 mins
- Added option switching relay for dial test
- optimized the test procedure of the dial test for reading of pulses more accuracy
Upgrade package program version up to 1.4.38 updated on 2022-12-6
The file GENY.Software.Update.1.4.38.EXE contains the following updated files
Metertest.exe, file version 1.4.205
DeviceControl.dll, file version 1.1.2.88
Yc189d_toolkit.exe, file version 1.1.3.22
Ctrcomm.dll, file version 2.2.43
EEM software fixed up and enhanced
fixed up and enhanced program "metertest.exe." with the latest file version to 1.4. 187
- Optimize the display of data results of the dial test, and the test can show the results
- Fixed the issue about the misspelling of the reference standard meter name "SZ_01_03F."
- fixed up French-language expression issue
- Optimized the function dial test and fixed up the issue about reading.
- added reactive test for 2-phase/3-wire meter
- fixed up an issue about communication in case of other brand power source with a K6D reference meter
- change the message box with parameters output in the manual operation of the power source
- optimized the dial test for less extra power
- Fixed up the issue about the error message of communication of radian meter: bad port handle
- allow users can input custom values of voltage & power factor to perform dial-test.
improved and fixed up toolkit program with the latest version number 1.1.3.20
- Setup current range for the precise output
- Added current loop switching function on the suspension test rack
- Added power consumption feature
- Optimize the dial test feature
- New feature VA test
- New feature ICT test
- New Starting test routine implement that the Time interval for the 1st Pulse & the 2nd Pulse can be displayed & recorded
along with the KWh reading, according to IS 13779:2020 Clause 12.14 - New Creep test routine implement that program can set any pulses then the display & record can be registered the Time-interval for the first two spurious pulse(s) along with the PASS/FAIL, according to IS 13779:2020 Clause 12.13,
improved and fixed up query data program "DataManage.exe" with the latest file version 1.1.0.37
- fixed up the error code "run-time error :13" as shown in the following screenshot
- fixed issue about exporting data to excel in user as administrator
file name: Rport_Format.xls
file name: Report_withInflu.xls
file name: Report_Format_PLN.xls
file name: Report_AIR-E.xls
file name: Report_Format_India.xls - Optimize the program for exporting test data to an EXCEL file
upgrade package program version 1.4.37
file name: GENY Software Update(1.4.37).EXE
EEM software fixed up and enhanced
Fixed issues for subprogram "DataManage.exe."
- Fixed issue #formatting of text output for pulsing error test
Fixed issues for main program "MeterTest.exe" with latest file version number 1.4.144
-
Fixed up the issue of output 1p3w / 2p3w / 2p3w +ab
-
Fixed up the issue of reactive power control of K6/K8 standard meters
Fixed issues for sub program "Yc189d_toolkit.exe", latest file version 1.1.3.16
-
Optimized the correction output for voltage/current, and the output accuracy can reach 100ppm, this process takes less than 1 minute.
-
Concurrent actions for multiple test systems
This feature enables multiple test benches to work simultaneously, and currently includes- communications testing
- power supply unit actions: frequency set, PSU output, PSU precise output, REF controlling
- Error calculator actions: pulsing test, readings
Take the parallel communication test of multiple PSUs as an example
- settings, input COM ports of PSUs like that
- run communication test, click the button to start, the returns will show on the log area, see the following screenshot for reference.
- stop the test, all working threads will be completed automatically and show the message "...finished", any other case to stop the test, just click the button to stop all working threads, please refer to below
Fixed issues for API library "Ctrcomm.dll" with file version number 2.2.41
- Fixed up the issue of reactive power control of K6/K8 standard meters
Full Changelog: 1.4.33...1.4.36
EEM version 1.4.35
Update list
- Added new reference standard meter SZ-01A-K3F
- Optimize the control process of standard power source YCSS-101
- Added new feature for High frequency pulsing test by Automatic frequency division
- fixed issue# superposition harmonic injection for type-d test bench
- Add pulsing error for dial test at toolkit program (ver. 1.1.3.11)
[Video reference]: https://v.youku.com/v_show/id_XNTgxMDUzNDYxNg==.html MD / Dail test with pulsing error - Fix up the 1P3W meter test problem
- Add new test for 2P3W electricity meter
- Optional working current for preheating
- Improved synchronization of multiple test benches
- Added Max demand test item in EEM sequence test program
Upgrade package & program files
GENY.Software.Update.1.4.35.EXE
Note: The latest dynamic link library “ctrcomm.dll” with file version 2.2.40 is included
Toolkit program (latest file version : 1.1.3.11)
Metertest test sequence program (latest file version : 1.4.139)
prjErrCalMgr.dll latest file version : 1.0.36
Full Changelog: 1.4.33...1.4.35
EEM version 1.4.34
Update list
- Added two updates excel files, test data export templates
- Fixed issue #system problems caused by calling different versions of Excel execution programs
- Fixed issue #data storage for influence test
- Support to the test of 1-phase 3-wire meters
- Fixed reading issues with creeping /starting /dial test for type-g test bench systems
- A large number of pulse/circles is allowed to input in the load point accuracy test
- Fixed reading issue for dial test on 3p3w mode
- Toolkit program for Harmonic and harmonic superposition output
- fixed issue #harmonic injection for type-g test bench
- Added new feature for calculation of test pulse number according to quantization error control rules
- Fixed issues on influence test for type-g bench
- Add new feature for dial test / MD test to utility program, and cumulative power and mean power to be displayed
Update files
- Metertest.exe
- Yc189d_toolkit.exe
- Calculation of the circle number of EEM quantization error priority.xlsx
Upgrade package
GENY.Software.Update.1.4.34.EXE